Automated clinical chemistry analyzers repeatedly perform sample aspiration, reagent dispensing, reaction mixing, and cuvette washing during routine operation. If any of these components is not cleaned properly, residual samples or reagents may enter subsequent tests and cause carryover or cross-contamination, affecting analytical results.

Carryover is an important issue in automated clinical chemistry testing. Common sources include insufficient cleaning of sample probes, reagent probes, mixers, and reaction cuvettes, as well as problems with the washing system.

Insufficient sample probe cleaning

The sample probe aspirates specimens from sample containers and transfers them into reaction cuvettes. If residual sample material remains on its inner or outer surface, it may be transferred into subsequent specimens during sampling and cause sample-to-sample carryover.

This problem can affect low-concentration samples following the measurement of a sample with a much higher analyte concentration, particularly when the probe washing process is inadequate.

How to address the problem:

Clean the inner and outer surfaces of the sample probe according to the instrument's maintenance procedure. Check for visible deposits and sample residues.

Inspect the washing solution supply, fluid lines, and rinsing components to confirm that the cleaning cycle is functioning properly.

Check the probe for blockage, bending, or damage. Clean or replace it according to the manufacturer's maintenance requirements when necessary.

When the analyzer supports special washing procedures, configure additional cleaning steps for test combinations that are prone to carryover.

Inadequate reagent probe cleaning

The reagent probe aspirates and dispenses reagents for different assays. If reagent residue remains inside or outside the probe, it may enter a subsequent reaction mixture and alter the reaction conditions, leading to inaccurate results.

Reagent carryover does not necessarily affect every combination of tests. Some reagent combinations are more sensitive to residual contamination than others.

How to address the problem:

Clean the reagent probe according to the manufacturer's procedure and inspect both its inner and outer surfaces for reagent residue.

Check the probe washing position, cleaning solution supply, and drainage system to identify any blockage or washing problems.

For assay combinations that are prone to reagent carryover, use the analyzer's dedicated anti-carryover settings or additional cleaning procedures where supported.

Inspect the reagent probe for wear or other mechanical problems. Contact qualified service personnel when replacement or further maintenance is required.

Insufficient mixer cleaning

The mixer helps combine samples and reagents to produce a uniform reaction mixture. If residual reaction liquid remains on its surface, it may enter another reaction mixture during subsequent tests and cause carryover.

After prolonged use, the mixer may also develop deposits, surface wear, or coating damage. These conditions should be checked during routine maintenance.

How to address the problem:

Clean the mixer according to the instrument's specified procedure, paying particular attention to surfaces that come into contact with the reaction mixture.

Inspect the mixer for deposits, scratches, and coating damage. Clean or replace the component according to the maintenance requirements.

Check the mixer washing position and operating condition to ensure that the cleaning cycle is completed properly.

Inadequate reaction cuvette cleaning

Reaction cuvettes hold the sample and reagents during chemical reactions and provide the optical measurement area. In analyzers that use reusable cuvettes, residual reaction liquid from a previous test may affect the absorbance or other measurement signals in subsequent tests.

Problems with cuvette washing may also originate from the wash nozzles, fluid lines, washing pumps, or drainage components. If the cleaning solution does not reach the cuvette properly or residual liquid is not completely removed, subsequent measurements may be affected.

How to address the problem:

Check the inside of the reaction cuvettes for visible residue, deposits, or other contamination.

Inspect the cuvette washing nozzles, fluid lines, and washing pumps to confirm that the cleaning solution is supplied and drained properly.

Run the cuvette washing procedure specified by the instrument manufacturer. Check the cleaning performance and condition of the cuvettes when necessary.

For analyzers that use disposable reaction cuvettes, replace them according to the manufacturer's instructions rather than treating them as reusable components.

Washing system malfunctions

Carryover may also result from problems with the washing system rather than contamination of individual components. Insufficient cleaning solution, blocked fluid lines, malfunctioning washing pumps, or restricted waste drainage can prevent the analyzer from completing its cleaning cycles properly.

How to address the problem:

Check the cleaning solution level, condition, and supply lines. Use cleaning solutions that meet the instrument manufacturer's requirements.

Inspect the washing pumps, rinsing nozzles, and fluid connections for blockages or irregular fluid delivery.

Check the waste lines and waste container to ensure that drainage is unobstructed and residual liquid does not accumulate in the system.

Review the washing program settings and confirm that the cleaning sequence, number of washing cycles, and related parameters comply with the manufacturer's specifications.

Preventive measures for reducing carryover

Preventing carryover requires routine cleaning, scheduled maintenance, and ongoing quality control. Cleaning components only after a problem occurs is not sufficient.

Proper sample and reagent handling

Samples and reagents should be stored and handled according to the instrument and reagent requirements. Avoid contamination of container surfaces, sampling areas, and storage locations. Check sample containers and reagent bottles before loading them into the analyzer, and address any visible contamination.

Routine cleaning and scheduled maintenance

Establish a maintenance schedule based on the instrument's operating manual. Clean the sample probes, reagent probes, mixers, reaction cuvettes, and washing components at the specified intervals. For test combinations that are prone to carryover, configure dedicated washing procedures when the analyzer supports this function.

Inspection of washing and fluid handling systems

Regularly check the cleaning solution supply, washing lines, pumps, and waste handling system. Cleaning agents and their use must comply with the manufacturer's specifications. Do not change the cleaning solution or adjust washing parameters without confirming that the proposed settings are suitable for the instrument.

Ongoing analytical quality control

After cleaning and maintenance, check the analyzer's operating condition and test performance according to the laboratory's quality control procedures. When abnormal results or shifts in quality control data occur, review the sample sequence, reagent usage, and washing records to investigate possible carryover.

Calibration should also be performed according to the instrument's requirements. However, calibration cannot replace cleaning when residual samples or reagents are causing contamination.

Carryover in automated clinical chemistry analyzers can involve liquid handling, mechanical components, and the washing system. Proper operating procedures, regular cleaning, and scheduled inspections help reduce the effects of residual samples and reagents on subsequent tests.

Custom precision machining for automated testing equipment

Automated clinical chemistry analyzers contain various mechanical components, including sample probe assemblies, mixer mechanisms, reaction cuvette supports, and other structural parts. Each component has specific dimensional and assembly requirements. For non-standard mechanical parts, we provide precision machining based on customer-supplied engineering drawings to meet the required structural and dimensional specifications.

Hansheng Automation also provides custom machining services for precision components used in other industries. Our work includes yacht components, filling machinery parts, remote-control racing car structures, and tobacco packaging machinery components. We manufacture non-standard parts according to customer drawings to support different equipment manufacturing and assembly requirements.