Clinical chemistry analyzers use optical and electrochemical methods to measure specific chemical components in blood, serum, plasma, urine, and other biological samples. Photometric analysis is one of the most commonly used measurement methods. Depending on the assay, analyzers use endpoint, kinetic, or fixed-time methods to calculate analyte concentrations or enzyme activities. Some models also include an ion-selective electrode (ISE) module for measuring electrolytes such as sodium, potassium, and chloride.

Basic principle of photometric analysis

Photometric analysis determines the concentration of a substance by measuring how much light a sample reaction mixture absorbs at a selected wavelength. A typical photometric system consists of six functional elements: a light source, a wavelength selector, a cuvette or reaction cell, a photodetector, a signal amplification and processing circuit, and a display and control unit. The optical arrangement varies between analyzer models. Some instruments use diffraction gratings or LED light sources instead of conventional filter-based systems.

The measurement process involves several stages.

Light source and wavelength selection

The light source provides the illumination required for measurement. A filter or diffraction grating selects a specific wavelength or narrow wavelength band. The selected wavelength depends on the optical properties of the reaction system used for each test.

Sample reaction and light absorption

The analyzer mixes the sample with the appropriate reagent in a cuvette or reaction cell. The chemical reaction may produce a change in color or turbidity. As light passes through the reaction mixture, part of it is absorbed or scattered.

Conversion of optical signals into electrical signals

The transmitted light reaches the photodetector, which converts the optical signal into an electrical signal. The signal processing system calculates absorbance from the incident and transmitted light intensities. It then uses the selected assay method and calibration data to determine the test result.

Result calculation and display

The analyzer runs the appropriate measurement program for each test. Depending on the assay, it records absorbance at specified time points or monitors absorbance changes throughout the reaction. The system calculates the analyte concentration or enzyme activity and displays and stores the result.

Absorbance (A) is defined as the base-10 logarithm of the ratio of incident light intensity to transmitted light intensity. Under suitable measurement conditions, the Beer-Lambert law relates absorbance to the molar absorptivity, optical path length, and concentration of the absorbing substance. Changes in path length or analyte concentration affect the measured absorbance. This relationship provides the basis for quantitative photometric analysis.

Common analytical methods used in clinical chemistry analyzers

Different tests use different reaction mechanisms and calculation procedures. Common methods used in photometric analysis include endpoint, kinetic, and fixed-time methods.

Endpoint method

The endpoint method measures the absorbance of a reaction mixture after the reaction has reached its specified endpoint or a stable stage. The analyzer calculates the result from the relationship between the measured signal and the analyte concentration.

Depending on the assay procedure, endpoint measurements may use single-wavelength or dual-wavelength photometry. The optical measurement may also use colorimetry or turbidimetry. Some procedures include measurements before and after the reaction to correct for reagent blanks or calculate the final analytical signal.

Endpoint methods are commonly used for tests such as total protein, albumin, glucose, and creatinine. They can also be used for selected specific protein assays and therapeutic drug monitoring tests, depending on the reagent formulation and analyzer capabilities.

Glucose oxidase should not be listed as a test parameter. It is an enzyme used as a reagent component in certain glucose assay methods.

Kinetic method

The kinetic method measures changes in the absorbance of a reaction mixture continuously or at defined time intervals under specified reaction conditions. The analyzer calculates the absorbance change per unit of time and uses the corresponding calibration relationship to determine the analyte concentration or enzyme activity.

For enzyme assays, the analyzer generally evaluates data collected during an appropriate linear phase of the reaction. By measuring changes associated with substrate consumption or product formation, the system can determine enzyme activity. Kinetic analysis is also used for the quantitative measurement of certain metabolites.

Common calculation approaches include continuous multi-point rate measurement. Reaction temperature, pH, reagent volume, and measurement timing must be controlled according to the assay requirements.

Fixed-time method

The fixed-time method measures the absorbance of a reaction mixture at two predetermined time points. The analyzer calculates the result from the change in absorbance over the specified interval. In some clinical chemistry analyzer manuals, this method is also called the two-point rate method or two-point kinetic method.

Unlike continuous kinetic monitoring, the fixed-time method uses measurements taken within a defined reaction interval rather than recording the entire reaction curve. It is suitable for assays in which the measured signal change has a quantitative relationship with the analyte concentration under the specified reaction conditions.

The fixed-time method is generally treated as a form of rate analysis. It should not be confused with the endpoint method, which measures the reaction after it has reached a stable endpoint.

Ion-selective electrode method

The ion-selective electrode (ISE) method measures changes in electrical potential associated with specific ions in a sample. Clinical chemistry analyzers equipped with an electrolyte module can use this method to measure potassium (K⁺), sodium (Na⁺), and chloride (Cl⁻).

During measurement, the ion-selective electrode uses a selective sensing membrane to respond to changes in the activity of the target ion. A reference electrode provides the measurement reference. The potential difference between the two electrodes changes according to the activity of the target ion in the sample.

The analyzer uses the Nernst relationship to relate the measured potential to ion activity. Calibration with standard solutions and the corresponding calculation procedure are then used to determine the ion concentration in the sample. This method is mainly used for clinical electrolyte testing.

Analytical methods and custom component machining

Clinical chemistry analyzers use photometric methods, kinetic analysis, and ion-selective electrode technology to measure different biochemical parameters. The selected method depends on the properties of the analyte, the reagent reaction mechanism, and the analyzer's configuration. Each measurement module requires compatible optical components, mechanical structures, and functional parts.

For non-standard mechanical components used in laboratory testing instruments and related automated equipment, we provide precision machining based on customer-supplied engineering drawings. Manufacturing is tailored to the component structure and specified machining requirements.

Hansheng Automation also provides custom machining services for other industries, including precision components for remote-control racing cars, yacht structural parts, tobacco packaging machinery, and filling machinery. We manufacture non-standard components according to customer drawings to support different equipment manufacturing and assembly requirements.

Laboratory--Analytical-Instruments