A conventional high-performance liquid chromatography (HPLC) system consists of five main functional units: the high-pressure solvent delivery system, sample injection system, separation system, detection system, and data processing and control system. These units work together to introduce the sample, separate its components, detect the resulting signals, and process the analytical data.
The module configuration, operating pressure, and detector options vary between HPLC manufacturers and models, but the basic analytical process remains the same. Understanding the function of each unit helps users evaluate instrument configurations and select equipment according to sample properties and analytical requirements.
High-pressure solvent delivery system
The high-pressure solvent delivery system stores, prepares, and delivers the mobile phase. Its main components include solvent reservoirs, a degassing unit, and a high-pressure pump.
Mobile phase reservoirs
Solvent reservoirs store the mobile phase used during analysis. Common materials include glass and polytetrafluoroethylene (PTFE). The appropriate material depends on the chemical properties of the solvent and the instrument connection design. Reservoirs should be kept clean to prevent contaminants from entering the mobile phase and affecting subsequent analysis.
Degassing unit
Dissolved air in the mobile phase can form bubbles when pressure or temperature changes. These bubbles may interfere with pump performance and detector signals. The degassing unit removes dissolved gases from the solvent. Common methods include online vacuum degassing and other suitable degassing techniques.
High-pressure pump
The pump delivers the mobile phase continuously to the injection system and chromatographic column at a specified flow rate. Flow accuracy, flow stability, and pressure resistance all affect chromatographic performance.
The operating pressure range varies between conventional HPLC pump models. Some systems support maximum operating pressures of up to 60 MPa. This is a reference specification for certain instruments, not a universal pressure requirement for all HPLC systems. Actual operating pressure also depends on column packing particle size, column length, mobile phase viscosity, and the selected flow rate.
Methods using gradient elution require additional solvent proportioning and mixing functions. These functions allow the system to adjust the composition of different solvents according to a programmed analytical method.
Sample injection system
The sample injection system introduces a specified volume of sample into the mobile phase so that it can enter the chromatographic column for separation. Common configurations include manual injection valves and automatic samplers.
Manual injection valve
A manual injection system commonly uses a six-port valve. By changing the valve position, the sample held in a sample loop is introduced into the mobile phase stream.
The sample loop determines the injection volume. Common capacities range from a few microliters to several tens or 100 μL, depending on the valve and loop configuration.
This arrangement has a relatively simple structure and is suitable for laboratories handling smaller sample batches or requiring manual control of the injection process.
Autosampler
An autosampler generally consists of a sample tray, injection needle, metering mechanism, injection valve, and control unit. It draws a specified sample volume according to a programmed sequence and transfers it into the mobile phase flow path.
An autosampler can process multiple samples continuously, reduce manual handling, and improve injection volume repeatability. It is useful in laboratories performing routine or batch analysis.
Important injection system performance parameters include injection volume accuracy, repeatability, and carryover control. These factors affect chromatographic peak consistency and quantitative results.
Separation system
The separation system mainly consists of the chromatographic column and column oven. The column performs the actual separation of sample components.
Chromatographic column
The column contains stationary-phase packing material. As the mobile phase carries the sample through the column, individual components interact with the stationary phase to different extents. These differences produce different retention times and allow the components to be separated.
Common analytical HPLC columns have internal diameters of approximately 2.1 to 4.6 mm and lengths of approximately 50 to 250 mm. These dimensions are typical reference ranges rather than mandatory specifications for all liquid chromatography columns.
Packing particle size, stationary-phase chemistry, column length, and internal diameter all influence separation efficiency, backpressure, and analysis time. Common stationary phases include C18, C8, phenyl, and amino phases. Selection depends on the chemical properties of the target compounds.
Column oven
The column oven houses the chromatographic column and maintains its operating temperature. Temperature changes affect mobile phase viscosity, analyte retention, and separation performance.
Stable temperature control helps reduce retention time variation and improves analytical repeatability. Some HPLC column ovens provide temperature control accuracy of ±0.1°C. The actual performance should be checked against the instrument manufacturer's specifications.
Detection system
The detection system is installed downstream of the chromatographic column. It measures the responses of the separated components as they elute and converts these responses into electrical signals for the data processing system.
Common HPLC detectors include UV-Vis detectors, diode array detectors, and fluorescence detectors. Each type operates on a different principle and is suitable for specific analytical applications.
Ultraviolet-visible detector (UV-Vis)
A UV-Vis detector measures changes in ultraviolet or visible light absorption at selected wavelengths. It is suitable for compounds with appropriate ultraviolet or visible absorption characteristics and is commonly used in conventional HPLC systems.
Some detectors cover a wavelength range of 190 to 800 nm. The actual range depends on the light source, optical design, and instrument model.
Diode array detector (DAD)
A diode array detector uses multiple photodiodes to collect optical signals at different wavelengths. It can record spectral information across a defined wavelength range during analysis.
Compared with single-wavelength or variable-wavelength UV detectors, a DAD provides additional spectral data. These data can be used to compare the UV absorption characteristics of chromatographic peaks and support peak purity assessment and compound identification.
Fluorescence detector (FLD)
A fluorescence detector measures the light emitted by sample components after excitation at a specified wavelength. It is suitable for compounds with natural fluorescence or substances that produce a fluorescence response after derivatization.
For compounds suitable for fluorescence detection, an FLD can provide high sensitivity and selectivity. Its application range is limited by the fluorescence properties of the target compounds, so it cannot replace UV detection for every analytical method.
Data processing and control system
The data processing and control system coordinates the operation of the HPLC modules and handles signal acquisition, chromatogram processing, and analytical calculations.
Instrument control
The control system uses dedicated software and hardware modules to manage the pump, autosampler, column oven, and detector. Operating parameters may include pump flow rate, gradient program, column temperature, sample sequence, and detection wavelength.
The individual modules operate according to the programmed method to maintain consistent sample injection, separation, and detection conditions.
Data acquisition and processing
After the detector converts the response of each component into an electrical signal, the data acquisition system records the signal and generates a chromatogram. The accompanying software can identify chromatographic peaks, perform peak integration, calculate peak areas, and process quantitative results.
Quantitative analysis generally requires a calibration relationship established using reference standards. The concentration of the target component is then calculated from the response of the corresponding sample peak.
The data processing system can also manage sample sequences, analytical methods, and test reports, supporting data retrieval and result traceability within the laboratory.
Coordination of the HPLC system components
The five functional units follow a defined flow path during analysis. The mobile phase is supplied by the solvent reservoir, passed through the degassing unit, and delivered by the pump. The injection system introduces the sample into the mobile phase, which carries it through the chromatographic column for separation. The separated components then enter the detector, and the resulting electrical signals are acquired and processed by the data system.
Pump flow stability, injection repeatability, column separation performance, and detector response all influence the final analytical results. Instrument configuration should therefore be selected according to the sample type, analytical method, and detection requirements.
Custom precision machining services
In addition to mechanical components used in laboratory and analytical instruments, we provide custom precision machining based on customer-supplied drawings. We handle both single-piece and small-batch orders. Hansheng Automation also provides machining services for precision yacht components, tobacco packaging machinery parts, filling equipment components, and RC racing car structural parts. Each component is manufactured according to its specified geometry and dimensional requirements.