Loading Capacity Limits for Pre‑Vacuum and Pulsating Vacuum Pressure Steam Sterilizers

Aug 24, 2026 Leave a message

Ryan Kim
Ryan Kim
Specializing in CNC machining services, Ryan ensures that all components are precision-machined to meet the highest standards. His expertise is crucial for maintaining the quality of our products.

For pre‑vacuum pressure steam sterilizers, the maximum loading capacity shall not exceed 90 % of the chamber volume. Exceeding this limit restricts steam circulation and penetration, which may compromise sterilization efficacy.

 

For pulsating vacuum pressure steam sterilizers, two conditions apply:

  • Upper limit: also 90 % of the chamber volume, for the same reason as above.
  • Lower limit: no less than 10 % of the chamber volume. Insufficient load (below 10 %) can cause the "small‑load effect," leading to uneven steam distribution and temperature fluctuations, which also negatively affect sterilization performance.

 