How to Troubleshoot a Universal Testing Machine Crosshead That Will Not Move Up or Down

Nov 22, 2026 Leave a message

Daniel Tan
Daniel Tan
Head of product marketing, Daniel is responsible for positioning Hansheng's products in the global market. His work includes developing marketing strategies and communicating our value proposition to customers worldwide.

 

The moving crosshead of a universal testing machine adjusts the test space and travels at a specified speed during tensile, compression, and other mechanical tests. If the crosshead cannot move, travels unevenly, or moves in the wrong direction, the cause may involve the mechanical transmission system, electrical controls, travel limit switches, or test software.

When a fault occurs, stop the test and make sure the machine is in a safe condition before troubleshooting. The possible causes and corrective actions vary depending on whether the machine uses an electromechanical or hydraulic drive system.

The crosshead moves slowly or gets stuck

If the crosshead can move but operates slowly, intermittently, or with abnormal resistance, inspect the mechanical transmission system first.

For electromechanical universal testing machines with screw drives, check the lubrication of the lead screw and nut assembly. Insufficient lubrication, accumulated debris, or mechanical binding can interfere with normal crosshead movement.

Also inspect the screw supports and the lower nut assembly. Check for loose locking screws and signs of abnormal wear in the transmission components.

If lubrication is insufficient or foreign material is present, clean and lubricate the components according to the machine maintenance manual. Any disassembly of the lead screw or nut should be performed by qualified service personnel to avoid affecting the alignment and accuracy of the transmission system.

For hydraulic universal testing machines, the inspection should focus on the hydraulic oil supply, actuator operation, and hydraulic valve assembly.

The crosshead does not respond when the up or down button is pressed

If the crosshead does not move after pressing the up or down button, check the power supply, machine operating status, and safety protection devices.

Common causes include:

Emergency stop activated: When the emergency stop is engaged, the drive system is normally disabled. Identify the cause of the shutdown and confirm that the machine is safe before resetting the emergency stop according to the operating manual.

Thermal overload protection activated: The crosshead drive may stop if the motor is overloaded or its protection circuit is triggered. A qualified electrical technician should identify and correct the cause before resetting the overload protection.

Travel limit switch activated or damaged: When the crosshead reaches its upper or lower travel limit, the control system may prevent further movement in that direction. Inspect the limit switches, mechanical stops, and associated wiring.

Machine not enabled: Some testing systems require the frame to be enabled through the control software or operating panel before movement commands can be executed.

Incorrect operating parameters: Check whether the crosshead speed in the test software has been set to zero and whether the system displays an overload or other interlock alarm.

If these checks do not resolve the problem, further inspection of the power circuit, control wiring, and drive system may be necessary. Work inside the electrical control cabinet should be performed by qualified personnel. <Cite refs={["turn857393search0","turn857393search2","turn857393search35"]}/>

Test results differ at different crosshead speeds

If the same material produces different test results at different speeds, first check whether the actual crosshead speed matches the programmed setting.

Material properties can be affected by strain rate. Therefore, even when the specimen and testing machine remain unchanged, different test speeds may produce different tensile strength, yield strength, or elongation results.

If the crosshead speed fluctuates noticeably during testing, inspect the following:

Check the programmed speed in the test software.

Verify that the actual crosshead speed matches the specified value.

Inspect the motor drive, screw transmission system, or hydraulic control system, depending on the machine design.

Verify the crosshead speed according to the machine requirements. Arrange for service if the measured speed falls outside the specified range.

The test speed or strain rate must follow the applicable testing standard to maintain consistent and comparable results.

The crosshead moves down but will not move up

If the crosshead moves downward normally but does not respond to the upward command, check the upper travel limit protection system.

Possible causes include an activated upper limit switch, a damaged switch, poor electrical connections, or an abnormal position signal detected by the control system.

First, check the current crosshead position and the status of the upper limit indicator. If the crosshead has reached its upper travel limit, use the normal control procedure specified in the operating manual to move it away from the limit position.

If the crosshead has not reached the upper limit but upward movement remains disabled, inspect the limit switch, wiring, and control circuit. Replace the switch if it is confirmed to be faulty and use a replacement that meets the equipment specifications.

Do not bypass the limit switch or disable the safety interlock to force the crosshead to move.

The crosshead moves in the opposite direction

If the crosshead moves downward when the up button is pressed, or upward when the down button is pressed, stop the operation immediately. Continued movement may cause a collision with the grips, specimen, or other machine components.

Possible causes include:

Incorrect movement direction parameters in the control software.

Abnormal motor drive or control signal configuration.

Incorrect motor wiring or phase sequence.

Directional control problems in the hydraulic valve assembly.

First, verify the operating command and software settings. Qualified service personnel should then inspect the electrical connections, motor drive, or hydraulic control circuit.

If the crosshead continues to move in the wrong direction after the relevant settings and connections have been checked, stop using the machine and contact the manufacturer for further diagnosis.

The crosshead shows noticeable deformation under load

A universal testing machine crosshead normally undergoes some elastic deformation when subjected to a load. This is part of the mechanical response of the load-bearing structure. However, the actual deformation must remain within the design requirements of the machine.

Noticeable bending, uneven deformation, or permanent deformation after unloading may indicate one or more of the following problems:

The test load exceeds the rated capacity of the machine.

The crosshead material or structural stiffness does not meet the design requirements.

Connections between the crosshead and columns are loose.

The loading position is offset from the designed load center, resulting in eccentric loading.

The crosshead or another load-bearing component has developed fatigue damage.

If abnormal deformation is observed, stop loading and inspect the machine. The manufacturer may need to evaluate the load frame, structural connections, and material condition before deciding whether repair or component replacement is necessary.

Elastic deformation under load alone does not indicate that the crosshead material is unsuitable.